ยตAUTOLABIII - Compact Design

The most basic in the Autolab family of instruments, the µAutolabIII in combination with the powerful NOVA software, can be used for all standard DC-electrochemical techniques. 

The instrument has a built in analog integrator, analog and digital inputs and analog and digital outputs, making it a versatile instrument despite its non-modular nature.
The analog input allows you to measure an external signal (e.g. voltage) and the analog output can be used to control Autolab accessories or external devices (e.g. Rotating Disk Electrode). The digital inputs and outputs can be used to control devices such as mercury drop electrodes. 

Key features:  
   
Electrode connections 2, 3
Potential range +/- 5 V
Compliance voltage +/- 12 V
Maximum current +/- 80 mA

Current ranges

10 mA to 10 nA

Potential accuracy +/- 0.2 %
Potential resolution 3 µV
Current accuracy +/- 0.2 %
Current resolution 0.0003 % (of current range)
Input impedance > 100 GOhm
Potentiostat bandwidth 500 kHz
Computer interface USB
Control software NOVA